INFO - PLT FED TTCrx OK (PLL ready, dll ready, frame synced) INFO - Set up PLT FED interface #0, base address = 0x44000000, ID = 34 INFO - Output will be stored in this file: /home/2017ColdBox/cmsplt/interface/calib/pltlevelsfast_calibration_output.txt INFO - PLTFED channel 0-1 (mFec 7-1 hubAddress 5 roc 1) INFO - Levels and [separations]: BLMax: 174.458 [21.4223], 1DC: 243.479 [30.2492], 2DC: 420.812 [20.515], 3DC: 658.667 INFO - PLTFED channel 0-2 (mFec 7-1 hubAddress 5 roc 2) INFO - Levels and [separations]: BLMax: 173.646 [15.8553], 1DC: 220.729 [23.0668], 2DC: 351.708 [22.1073], 3DC: 542.042 INFO - PLTFED channel 0-3 (mFec 7-1 hubAddress 5 roc 3) INFO - Levels and [separations]: BLMax: 174.646 [0.093521], 1DC: 174.875 [0.945828], 2DC: 177.167 [0.956009], 3DC: 179.583 INFO - Beginning fine timing calibration... ERROR - Error: ot all fastors are landing in the same bucket. Either that or there are no fastors!!! INFO - MaxBucket after adjustment for pltFedId 0 pltFedChannel 1: 10 INFO - MaxBucket after adjustment for pltFedId 0 pltFedChannel 2: 10 INFO - MaxBucket after adjustment for pltFedId 0 pltFedChannel 3: 15 WARN - Something looks not quite right so: writing params files to calib/ WARN - If you want these settings you have to copy them by hand ERROR: couldn't find conf/params_pltfed34.dat; perhaps the calibration failed? Files will not be moved. ERROR: pltfed1_id is not set in calibration-configuration.conf; file will not be moved